GDYD-53A Automatic AC/DC Hipot Test Set
• 10inch color LCD screen, with PLC automation control.
• HV Voltage, HV current, LV voltage, LV current measured at the same time, with high precision sensor and high performance collection chip.
• The human-machine interface adopts full touch and partial key operation mode, intelligent working process, optional automatic mode and manual mode.
• Display HV voltage, HV current, LV voltage, LV current, time and withstanding result.
• Over-voltage, over-current and flashover protection. Output voltage, HV protection current, LV protection current (hardware has been set), mA meter protection current and timing time can be set.
• Support stage boosting (5 stages), and the voltage value of each stage can be arbitrarily set to facilitate the detection of test object problems.
• Zero-starting function. Test can be proceed after the voltage is back to zero.
• Store up to 8000pcs data in 3months.
• When setting voltage is arrived, it will timing automatically. After timing, voltage will be back to zero.
• When the LV current is over setting value, the voltage will be cut off automatically and it will be back to zero. When the HV current is over setting value, the test object will be cut off and test can be proceed continuously.
• With anti-interference ability, suitable for electromagnetic environment.
• Automatic error diagnosis, easy to find and solve problems, simple and convenient maintenance of equipment.
Control Unit
• Input voltage: AC220V±10% 50Hz±1
• Output capacity: 3kVA
• Low voltage output: AC 0~250V
• Output current: 15A (Max)
• Low voltage voltage measurement range: 0-300V resolution: 0.1V
• Low-voltage current measurement range: 0.2-15A Resolution: 0.1A
• High voltage measurement range: AC 0~50kV, DC 0~70kV Resolution: 0.01KV
• High voltage current measurement range: AC 1.0~60mA, DC 1.0~42.8mA Resolution: 0.01mA
• Milliamp meter current measurement range: 0 ~ 20mA Resolution: 0.01mA
• Timing range: 0~999s
• Voltage accuracy ≤±3.0%FS
• Current accuracy ≤±3.0%FS
Transformer
• Output capacity: 3kVA
• Low voltage (input) voltage: AC 0~200V
• Low voltage (input) current: AC 0~15A
• High voltage (output) voltage: AC 0~50kV, DC 0~70kV
• High voltage (output) current: AC 0-60mA, DC 0~42.8mA
• Measurement ratio: 500:1 Error: ±1%
• Impedance voltage: ≤12%;
• No-load current: ≤10%;
• Running time: 5min open at rated capacity, 5min off continuous cycle duty system, long-term operation at 1/3 rated voltage and 1/3 rated current.









